Temperature Range | 10ºC to 60ºC |
Humidity Range | Upto 80 RH |
ICH Conditions | • 25ºC / 65% RH • 30ºC / 65% RH • 40ºC / 75% RH • 30ºC / 75% RH |
Accuracy | ± 0.5 0C and 2% RH |
Accuracy | ± 0.5 0C and 2% RH |
Uniformity | ± 1.0 0C and 3% RH |
Controller System | Microprocessor Based PID Controller |
Dry Sensor | RTD Type (Pt-100) |
Wet Sensor | Humidity capacitance 3.3v volt |
Memory | 6000 Log Memory with Printing Interface (Data logger, Printing Data shows set Value, Actual, Value Date, Month, Year & Time) |
Inside Material Outside Material |
Stainless Steel 304 / 22 S.W.G Stainless Steel 304 / 20 S.W.G |
Body & Door Insulation | CFC Free Poly Urethane Foam (PUF) 75 mm Thickness Viewing Window Full View Acrylic Observation Door, Chamber Illumination Fluorescent Tube is provided in the Chamber. Port Hole 50 mm dia Porthole is provided on R.H.S of the Chamber for Validation |
Heating System | S.S Nichrome Wire Emended ‘U’ Shape Air Heater. Capacity - 1 Kw (depend upon Chamber size ) |
Cooling System | Cooling System Consists of Condenser & Compressor, Capillary Type By – Pass System |
Compressor | Hermetically Sealed CFC Free Compressor, Make: Kirlosaker Copeland |
Controller System | Temperature & RH% is controlled by Microprocessor Based PID Controller with Printer Interface Facility & PT-100 Sensor for Dry & RH % Capacitor Type Sensor for Wet Printing Data Shows Set Value, Actual Value Date, Month, Year & Time. |
- For boiler safety system For temperature safety system .
SAFETY FEATURES :
SAFETY FEATURES :
- Our safety controller is having three safety cut off :
- Lower Temperature Cut- Off ( Above Set Point )
- Lower Temperature Cut- Off ( Above Set Point )
- Lower Temperature cut-off ( Below Set Point ):
- The LED (LO1) blinks when temperature is undershoot below the lower set
Point (Lower cut-off : Below set point) - The LED (LO2) blinks when temperature is overshoot above the set Point (Lower
cut-off : Above set point - The LED (HT) blinks when temperature is overshoot above the set Point (Higher
Cut-off : Above set point) - The LED (BT) blinks when water is not present in the boiler tank.
- Microprocessor based Safety Controller (Auto Reset) which will be cut off the heater Output Supply in case of over shoot of Low Set Temperature and it gives overshoot indication and alarm.
- One Pt-100 Sensor Fitted inside the chamber. This will show the actual temperature on Safety Controller.
FOR BOILER SAFETY SYSTEM :-
- Thermostat is fitted in the boiler tank. Boiler Safety Circuit will cut off the chamber output supply in case of Low Water Level and it gives overshoot indication and alarm.
FOR TEMPERATURE SAFETY SYSTEM :-
- • Two Safety set points are available. One is lower set point and another is Higher set point.
- • If temperature gets Over Shoot above the Higher Set Point, Higher Safety will be cut off the chamber Power Supply.
- • If temperature gets overshoot above the lower set point , Higher safety will be cut off the heater Supply.
- • If temperature gets Over Shoot above the Higher Set Point, Higher Safety will be cut off the chamber Power Supply. Work on 220/230 Volts A.C.
MODEL NO | INNER CHAMBER SIZE | OUTER MILD STEEL POWDER COATED & INNER SS 304 | STAINLESS STEEL 304 |
---|---|---|---|
SSI-104 | 200 LITERS | Rs. /- each | Rs. /- each |